The EH8300 hematology analyzer of Guangzhou Excbio Technology Co., Ltd. has a linear range of PLT: 0-1000 x 109/L. In order to answer the customer's doubts that the platelet value of dengue fever is low, we do low-value platelet linear test.
According to the test results and data analysis, the EH8300 hematology analyzer of Guangzhou Excbio Biotechnology Co., Ltd. has a linearity and accuracy of PLT in the range of less than 50 x 109 / L, which can accurately test the low value of dengue PLT.
As a result, customer needs are met.
At the same time, it can be seen from the data that the linearity of parameters such as WBC, RBC, HGB and HCT is also very good。
Details of the experiment are in the attachment.
A report on low-value PLT linear tests for simulated dengue fever testing by EH8300